In principle, yes, but you may not be able to use it if there are detergents in your cell lysis buffer. For example SDS would interefere with the assay at relatively low concentrations.
You could do a simple check by doing a calibration graph using the lysis buffer to make the BSA dilutions, then adding to the assay as normal. Is always best to dilute the standard in the same buffer as the sample to check for interference or effects on linearity.
BSA is normally used as the standard in most protein assays. You just have to be aware that you are measuring protein concentration in BSA equivalents.
We used to use the Lowry and Bradford assays but now routinely use the BCA assay as it is less prone to interference and relatively sensitive, and the reagent goes for the peptide backbone rather than specific groups as in the case of the other assays. However, we always dilute the standard in the same buffer as the sample for the calibration graph.
In principle, yes, but you may not be able to use it if there are detergents in your cell lysis buffer. For example SDS would interefere with the assay at relatively low concentrations.
You could do a simple check by doing a calibration graph using the lysis buffer to make the BSA dilutions, then adding to the assay as normal. Is always best to dilute the standard in the same buffer as the sample to check for interference or effects on linearity.
BSA is normally used as the standard in most protein assays. You just have to be aware that you are measuring protein concentration in BSA equivalents.
We used to use the Lowry and Bradford assays but now routinely use the BCA assay as it is less prone to interference and relatively sensitive, and the reagent goes for the peptide backbone rather than specific groups as in the case of the other assays. However, we always dilute the standard in the same buffer as the sample for the calibration graph.
I've just managed with my measurements. In our case cell-lysate was get by freeze/thaw, so there were no detergents. Though the information about detergents influence was new for me. I do believe it will help me in future experiments as well as a recommendation about BCA assay.
Samples and standards should always be diluted in the same buffer and the same volumes added in the assay to minimise risks of interferance.
However, before assaying samples that may be available in very limited amounts, you should always run a 'normal' assay in parallel with a set of standards diluted with the sample buffer to ensure that it doesnt affect the absorbance readings or that at least it gives good linearity