Depends on the conductivity of your sample! Please, make a pellet (about 13 mm or higher of diameter per 1 mm thickness), paint Ag colloidal on the parallel surfaces and measure the conductivity. If the conductivity is small, you need measure a four-probe measurement. If not, you can measure the in a two-probe measurement. Observe, when you are painting the surfaces, please paint one side full and the other side you make a bullet contact. So, if the conductivity is small, you can make the other contacts and can use the same sample. When the conductivity is medium (at around 10^6) perhaps you can use a conventional 51/2 digits voltmeter and a DC voltage source, that are simple instruments. Remember, to measure the DC conductivity you need to measure the resistance, which is a size-dependent property. Thus, you can make a large sample in diameter and small in thickness in order to use the above instruments. Also, if your sample is a ceramic one, it's better, always, measure the resistance between two faces. In this form you observe the pores contribution.
You should prepare flat specimen (e.g. a disk) with a diameter of few cm-s and a thickness of 1-2 mm. Then you should paint or evaporate concentric metal electrodes on both sides (you may use a guard ring on one side too) and you can try to measure the conductivity by a simple ohmmeter. (Depending on the conductivity the test voltage and the sensitivity of the ohmmeter should be changed). Under DC conditions you get a rough estimate. Of course, using a broadband spectrometer you can get a more complex picture at several frequencies.
Nathapong, Im in Thailand ! , come to me, meaning send me your samples, then your problem will be solved. . anyway, please keep in mind those useful techniques the other researchers answered you .
Four-probe method can be used for determination of DC conductivity. See for example, J. Electrochem. Soc. 1996 143(5): 1736-1744; doi:10.1149/1.1836709. Or two-probe method can be used for impedance spectroscopy analysis.
Depends on the conductivity of your sample! Please, make a pellet (about 13 mm or higher of diameter per 1 mm thickness), paint Ag colloidal on the parallel surfaces and measure the conductivity. If the conductivity is small, you need measure a four-probe measurement. If not, you can measure the in a two-probe measurement. Observe, when you are painting the surfaces, please paint one side full and the other side you make a bullet contact. So, if the conductivity is small, you can make the other contacts and can use the same sample. When the conductivity is medium (at around 10^6) perhaps you can use a conventional 51/2 digits voltmeter and a DC voltage source, that are simple instruments. Remember, to measure the DC conductivity you need to measure the resistance, which is a size-dependent property. Thus, you can make a large sample in diameter and small in thickness in order to use the above instruments. Also, if your sample is a ceramic one, it's better, always, measure the resistance between two faces. In this form you observe the pores contribution.