Heat it to 280 °C or higher (depending from the max. temperature of your column brand) for at least 30 min and purge with mobile phase. Inject solvent and heat again until the impurities disappear.
Usually cut 1-2 meteres of front end helps in reducing contamination. There are some kinds of backwards ellution with solvents but its tricky and not so much helpfull.