High vacuum is not measured and the system shows the error "0d9f: filament in ion gauge consumed". Both filaments in ion source have been changed by new ones and the ion sorce have been cleaned. What could be happening?
"filament in ion gauge consumed" is not indicating the filament in the MS source. It is the "high vacuum reader" and probably should be replaced by Shimadzu service. You can try to see if there are any big leaks in the system.