02 February 2016 8 9K Report

Dear all,

I am doing single-edge-notched-bending (SENB) fracture-toughness tests on brittle epoxy samples according to ISO 13586. Now I was wondering how the GIc measured this way agrees with the theoretical definition of the GIc. In an SENB test, the GIc is basically the mechanical energy at max. load (i.e., at fracture), devided by the residual cross-secion area of the specimen. The definition of GIc says that it is the energy dissipated per area, so in this example this calculation is only valid if the material breaks all the way through and the whole mechancial energy is dissipated that way.

Now, the crack often does not propagate all the way through the specimen. In such a case, the created fracture surface is actually smaller than the cross-section area that is used in the calculation.

Likewise, the force might not drop to zero upon fracture. In such a case we use a too high mechanical energy in the calculation, as there was still some energy stored after the fracture event.

Finally, I observed that some specimens broke all the way though and then jumped off the measurement table upon fracture, i.e., there was still plenty of mechanical energy stored in the material, which would have sufficed to propagate the crack much further. So in this case we use a too high mechanical energy as well.

Is the GIc that we measure in an SENB test really the correct value from a scientific point of view?

Yours sincerely, Dietmar Haba

Similar questions and discussions